Verification, Equipment required – Agilent Technologies System DC N5700 User Manual

Page 94

Advertising
background image

Appendix B Verification and Calibration

94

Series N5700 User’s Guide

Verification

Verification procedures verify that the power supply is operating
normally and is within published specifications. There are two types
of verification tests:

Performance

These tests verify that the power supply meets all of the specifications listed
in Appendix A. They can also be used to verify that the power supply is
properly calibrated.

Calibration

These procedures calibrate the power supply.

If the power supply fails any of the tests or if abnormal test results
are obtained, try calibrating the unit. If calibration is unsuccessful,
return the unit to an Agilent Technologies repair facility (see
Appendix D).

Equipment Required

The equipment listed in the following table, or the equivalent to this
equipment, is required for the calibration and performance tests. A
test record sheet may be found at the back of this section.

Type Specifications

Recommended

Model

Digital Voltmeter

Resolution: 10 nV @ 1V;
Readout: 8 1/2 digits
Accuracy: 20 ppm

Agilent 3458A or equivalent

Current Monitor

15 A (0.1Ω) 0.04%, TC=5ppm/°C

100 A (0.01Ω) 0.04%, TC=5ppm/°C

300 A (0.001Ω) 0.04%, TC=5ppm/°C

Guildline 9230/15
Guildline 9230/100
Guildline 9230/300

Load Resistor

For 750 W models:
0.06Ω, 0.09Ω, 0.21Ω, 0.53Ω, 1.20Ω, 2.11Ω, 4.80Ω, 8.42Ω,

13.33Ω, 30.0Ω, 120Ω, 462Ω - all resistors 1kW minimum.

For 1500 W models:
0.03Ω, 0.04Ω, 0.10Ω, 0.26Ω, 0.60Ω, 1.05Ω, 2.40Ω, 4.21Ω,

6.67Ω, 15.0Ω, 60Ω, 231Ω - all resistors 2kW minimum.

Electronic Load

150 V, 100 A minimum (for Models N5741- N5750A)
150 V, 180 A minimum (for Models N5761- N5770A)

Agilent N3300A mainframe,
with 3 - N3305A modules

GPIB Controller

Full GPIB capabilities (for calibrating the unit over the GPIB) Agilent 82350B or

equivalent

Oscilloscope

Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip

Agilent Infiniium or
equivalent

RMS Voltmeter

True RMS
Bandwidth: 20 MHz
Sensitivity: 100 µV

Rhode and Schwartz Model
URE3 or equivalent

Differential Amplifier

Bandwidth: 20 MHz

LaCroy 1855A, DA1850A,
or equivalent

Terminations

1 – 50Ω BNC termination

2 – 50Ω, 1/8 W termination resistors

Variable-voltage xfmr
or AC source

Adjustable to highest rated input voltage range.
Power: 2000 VA

Agilent 6813B or equivalent

Advertising