Quick Reference 1

Series N5700 User’s Guide

13

J2 Sense Connector

1 – Remote sense (+)
2 – Local sense (+)
3 – Not used
4 – Local sense (–)
5 – Remote sense (–)

The factory-shipped configuration is shown in the figure.

SW1 Setup Switch

The factory-shipped setting is Down for all switches.

1 – Output voltage, voltage
programming

Down: The output voltage is programmed by the front panel.
Up: The output voltage is programmed by the external voltage signal.

2 – Output current, voltage
programming

Down: The output current is programmed by the front panel.
Up: The output current is programmed by the external voltage signal.

3 – Programming range
(voltage/resistance)

Down: The remote programming range is: 0 – 5V / 0 – 5KΩ.
Up: The remote programming range is: 0 – 10V / 0 – 10KΩ.

4 – Voltage and Current
monitoring range

Down: The remote monitoring range is: 0 – 5V.
Up: The remote monitoring range is: 0 – 10V.

5 – Shut-Off Logic Select

Down: OUT OFF = Low (0 – 0.6V) or short; OUT ON = High (2V – 15V) or open.
Up: OUT OFF = High (2V – 15V) or open; OUT ON = Low (0 – 0.6V) or short.

6 – Not Used

7 – Output voltage, resistive
programming

Down: The output voltage is programmed by the front panel.
Up: The output voltage is programmed by the external resistor.

8 – Output current, resistive
programming

Down: The output current is programmed by the front panel.
Up: The output current is programmed by the external resistor.

9 – Enable/Disable control

Down: The J1 Enable+/Enable– pins are not active.
Up: The J1 Enable+/Enable– pins are active.

1

2

3

4

5

6 7

8

9