2 Installation

18

Series N5700 User’s Guide

Rack Installation

CAUTION

Ensure that the screws used to attach the rack slide kit do not penetrate more
than 6 mm into the sides of the unit.

Do not block the air intake at the front, or the exhaust at the rear of the unit.

The Agilent N5700 power supplies can be mounted in a standard 19-
inch rack panel or cabinet. They are designed to fit in one rack unit
(1U) of space. To install the power supply in a rack:

1. Use the front panel rack-mount brackets to install the power
supply in the rack.

2. Use a support bracket to provide adequate support for the rear of
the power supply.

3. If using rack mount slides, use Agilent N5740A Rack-mount Slide
Kit

to install the unit in a standard 19-inch equipment rack. Refer to

the following figure for assembly instructions. Use two #10-32 x 3/8
in (max.) screws at each side. To prevent internal damage, use the
specified screw length only.

Cleaning

WARNING

SHOCK HAZARD To prevent electric shock, unplug the unit before cleaning.

Use a dry cloth or one slightly dampened with water to clean the
external case parts. Do not attempt to clean internally.