Zero Adjustment Using Display Unit Switches

This section describes the procedure for zero adjustment using the display unit switches.
(For more details regarding setting methods using these switches, refer to Chapter 5: Basic
Operating Procedures.)

The parameters for zero adjustment are B50/M10: Auto Zero Exe

(and either of these

can be used to carry out this procedure). For more details regarding these parameters, refer
to Chapter 6: Parameter Description.

The parameter M10: Auto Zero Exe

will be used in the following description.

Once in Setting Mode, use the

switch to

move the cursor to M: Ajustment

.

Touch the

switch to access Sub-item

Parameter Search Mode.

Upon selection of M: Ajustment, the cursor will be
positioned at M10: Auto Zero Exe

. (Sub-item

sellection (D))

Touch the

switch to access Parameter

Replecement Mode.

Touch the

switch to move the cursor to

“Execution”.

Touch the

switch to select “Execution”.

In order to request confirmation, the entire display

flashes on and off. Touch the

switch once

again at this time to fix selection of the automatic
zero adjustment function.

Auto zero adjustment function is being executed
(about 30 seconds).

When zero adjustment function has been completed,
the system returns automatically to the sub-item
selection screen (D).

F0901.EPS

Setting Mode

Major item

parameter

Setting Mode

K:Diagnosis

M:Adjustment

N:Test

M10:Auto Zero Exe
No Execution

No Execution

Execution

M10:Auto Zero Exe
No Execution

Execution

No Execution

Major item

Sub- item

selection (D)

Major item

Sub- item

selection (D)

M:Adjustment

10:Auto Zero Exe

11:Magflow Zero

M:Adjustment

10:Auto Zero Exe

11:Magflow Zero

M10:Auto Zero Exe

Execution

Sub-item

Parameter

Search Mode

Parameter

Replacement

(execute) Mode

Sub-item

Parameter

Search Mode

Start:

Major Item

Parameter

Search Mode

Now Auto Zero

Executing...