8. OPERATION VIA HART COMMUNICATOR

HART Communicator
Polling
Never Poll
Never Poll
Ask Before Polling
Always Poll
Digital Poll

HELP

SAVE

ESC

ENTER

6

Select “Digital Poll” and press ENTER (F4).

F4

(ENTER)

×

3

5. HART Output

Online Menu

Main Menu

4. Utility

1. Configure Communication

1. Polling

AXFA11 : FI-1201
HART Output
1 Poll addr 1
2 Num req preams 5
3 Burst mode On
4 Burst option PV

HELP

SAVE

HOME

ENTER

1

2

Return to the “Online” menu with HOME (F3).

Return to the “Main” menu with a “previous” key.

Select “Utility”.

Select “Configure Communication”.

Select “Polling”.

F3

(HOME)

Open the “Auto Poll” display:

AXFA11 : FI-1201
Online
1 Device Setup
2 PV

20m3/h

3 PV AO1

20mA

4 PV Span

20m3/h

HELP

SAVE

HOME

ENTER

3

JLK

4

HART Communicator
1 Offline
2 Online
3 Frequency Device
4 Utility

HELP

SEND

HOME

ENTER

4

HART Communicator
Utility
1 Configure Communic
2 System Information
3 Listen for PC
4 Storage Location
5 Simulation

HELP

SEND

HOME

ENTER

5

HART Communicator
Configure Communica
1 Polling
2 Contrast
3 Off Time
4 Ignore diagnostics
5 Delete Configs

HELP

SEND

HOME

ENTER

F0819.EPS

F3 : HOME

STU

1

or

STU

1

or

NOTE

1. If “Never Poll” is selected in “Polling” when the address is assigned, the

“Online Menu” cannot be called up and displayed. Be sure to select “Digital
Poll” in “Polling” after assigned the polling address.

2. When the same polling address is assigned for two or more field devices in

multidrop mode, communication with these field devices is disabled.