IM 01E20C01-01E

6-5

6. PARAMETER DESCRIPTION

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

T0604-2.EPS

C60 —

(SELF CHECK)

See “6.5 Alarm Functions”.

Good
Error

R

m/s

3

C44 Velocity Check

(VELOCITY CHK)

Display of the span setting using
flow velocity (m/s).

0.000 to 99.999

R

C45 Density Unit

(DENSITY UNIT)

Sets units for density when mass
flow rate is selected.

kg/m

3

lb/gal
lb/cf

W

kg/m

3

C45

0 to 4

C46 Mass Flow Density

(MASS DENSITY)

Sets density when mass flow rate
is selected (with units from C45).

0 to 32000

W

0

C47 User Span Select

(USER SPN SEL)

Selects whether or not special
units are used for flow rate units.

No
Yes

W

No

C48 Flow User Unit

(FL USER UNIT)

Sets the special flow rate units.

8 alphanumeric
characters

W

C48

0 to 4

C49 Flow User Span

(FL USER SPAN)

Sets span when using special flow
rate units.

0.0001 to 32000

W

100

(4) Item D (Menu D): Total Setting items

Menu D contains setting items such as the totalization scale and the forward/reverse totalized values.

T0605.EPS

Item

Name

Display unit

(BRAIN)

Data range

Display unit

/BRAIN

Default value

(*): Indicated item

Units

R/W

Position

of decimal

point

Description

D10 Total Unit

(TOTAL UNIT)

W

n Unit/P
u Unit/P
m Unit/P
Unit/P
k Unit/P
M Unit/P
Pulse/s

Pulse/s (*)

Selects the flow rate unit per one
pulse as used for totalization
display. Linked with B30.

D12 Total Decimal Pnt

(TL DECIMAL)

W

0
1
2
3
4
5
6
7

0

Selects position of decimal point
for totalization display

D11 Total Scale

(TOTAL SCALE)

W

0 to 32000

D10

(B30)

0 to 4

0 (*)

Sets the flow rate per one pulse
for the totalization display.
Linked with B31.

Sets the range in vicinity of 0%
within which the totalization
display will be halted.

D13 Total Low Cut

(TOTAL LOWCUT)

W

0 to 100

%

0

3 %

Sets the totalization preset value
in the lower 6 digits of the 8-digit
totalized value.

D21 Ttl Set Val Lower

(TL SET VAL L)

W

0 to 999999

0

0

Sets the totalization preset value
in the upper 2 digits of the 8-digit
totalized value.

D22 Ttl Set Val Upper

(TL SET VAL U)

W

0 to 99

0

0

Sets the totalization switch value
in the lower 6 digits of the 8-digit
totalized value.

D23 Ttl Switch Lower

(TL SWITCH LO)

W

0 to 999999

0

0

Sets the totalization switch value
in the upper 2 digits of the 8-
digits totalized value.

D24 Ttl Switch Upper

(TL SWITCH UP)

W

0 to 99

0

0

Selects whether or not special
units are used as totalized units.

D30 Ttl User Select

(TL USER SEL)

W

No
Yes

No

Sets the special totalized units.

D31 Ttl User Unit

(TL USER UNIT)

W

8 alphanumeric
characters

See “6.5 Alarm Functions”.

D60 —

(SELF CHECK)

R

Good
Error

Executes “Start” or “Stop” of the
totalization function, or executes
“Preset Total” or “Preset Rev
Total”.

D20 Total Execution

(TOTAL EXEC)

W

Start
Stop
Preset Total
Preset Rev Total

Start

D00 Total Set

(TOTAL SET)