IM 01E20C01-01E

8-10

8. OPERATION VIA HART COMMUNICATOR

8.3.4.5

Example 2 of Other Settings: Power Frequency (For DC version only)

IMPORTANT

In situation where a DC power supply is used for converters, set the local com-
mercial power frequency in area where the converter is lnstalled. The flowmeter is
set to 50.00 Hz at the factory.
Set “No” for “Power Synch” and the local commercial power frequency for
Power Frequency”.

1. Device Setup

4. Detailed Setup

4. Function Set

5. Aux

Power Frequency

AXFA11 : FI-1201
Power Frequency
50.00 Hz
50.00

DEL

DEL

ESC

ENTER

2

3

Enter “60.00” with alphanumeric keys.

Press ENTER (F4).

Press SEND (F2) to set the data into the AXFA11.

Press HOME (F3), and return to the “Online” menu.

4

Open the “Power freq” setting display:

‘60.0’

AXFA11 : FI-1201
Power Frequency
50.00
60.00

DEL

DEL

ESC

ENTER

F4

(ENTER)

F2

(SEND)

F3

(HOME)

F0813.EPS

AXFA11 : FI-1201
Aux
6 Dead Time Os
7 Pulsing Flow No
8 T/P Damp Select
9 Power Synch No
Power Frequency

HELP

SAVE

HOME

ENTER

1

AXFA11 : FI-1201
Aux
6 Dead Time Os
7 Pulsing Flow No
8 T/P Damp Select
9 Power Synch No
Power Frequency

HELP

SAVE

HOME

ENTER

Select “Power Frequency”.

Ч

9