Setting

Function

Halt the adhesion diagnostic function

No

Yes

T0642.EPS

Carry out the adhesion diagnostic function

NOTE

Adhesion Diagnostic Function

᭿

This function diagnose adhesion using elec-
trode resistance values.

᭿

When “Adhesion check” has been set for B41/
H11: Display Select 2
or B42/H12: Display
Select 3
, the diagnose adhesion is indicated
on the display unit using four different levels.

᭿

If the judgment value for Level 3 is exceeded,
a warning is displayed; and if the value for
Level 4 is exceeded, an alarm is displayed.

᭿

Available conductivity for this function is
limited to:
Nominal size 10 mm or smaller: 30

S/cm

Nominal size 15 mm or larger: 10

S/cm

Make sure to use the adhesion diagnostic
function with the greater conductivity than the
above mentioned value.

F0619.EPS

Alarm

Level: 4

Warning

Level: 3

Level: 2

Level: 1

[K11: Adhesion Level1]

Setting the resistance value

for adhesion level 1.
This parameter sets the resistance value (in M ohm) for
judgment of Level 1.

[K12: Adhesion Level2]

Setting the resistnce value

for adhesion diagnostic level 2.
This parameter sets the resistance value (in M ohm) for
judgment of Level 2.

[K13: Adhesion Level3]

Setting the resistance value

for adhesion diagnostic level 3.
This parameter sets the resistance value (in M ohm) for
judgment of Level 3.
*: The warning

80: Adhesion Wng

is displayed

when the adhension level reaches Level 3.

*: If “Warning Output” has been selected for

F10:

SO1 Function

or

F11: SO2 Function

, then

status output will be performed when the adhesion
level reaches Level 3.

[K14: Adhesion Level4]

Setting the resistance value

for adhesion diagnostic level 4.
This parameter sets the resistance value (in M

ohm) for

judgment of Level 4.
*: The process alarm

33: Adhesion Alm

is displayed

when the adhension level reaches Level 4.

*: Alarm output will be performed if “Yes” has been

selected for

G34: Alm-Adhesion

.

[K15: Adh Measure Value]

Displays the resistance

value for adhesion diagnose
This parameter displays the value measured using the
adhesion diagnostic function (in M

ohm).

When “No” is selected for

K10: Adhesion Check

,

this parameter displays the indetermination value.

(10) Menu M: Automatic Zero Adjustment

Function Setting items

Menu M contains items that are relevant to automatic
adjustment of the zero point.

[M10: Auto Zero Exe]

Execution of automatic zero

adjustment function

Refer to the description of parameter

B50

.

[M11: Magflow Zero]

Results of automatic zero

adjustment
This parameter is used to display the results obtained
from

B50/M10: Auto Zero Exe

. Specifically, the

correction value displayed, and it is also possible to
directly enter correction value.

(11) Menu N: Loop Test Setting items

Menu N contains items that are relevant to loop testing.

[N10: Test Mode]

Setting for loop test execution

Setting

Function

No execution of loop testing

Loop testing is started

Normal

Test

T0644.EPS

IMPORTANT

(1) Test output has priority over flow rate mea-

surement signals. When carrying out flow
rate measurements, be sure to always return
to “Normal”.

(2) Upon entry to the Test Mode, all output

terminals will simultaneously adopt test
condition.