Example 2: for adjustments using a voltmeter, Select the “ scaled d/a trim ”. press ok (f4) – APC AXFA11G User Manual

Page 87

Advertising
background image

IM 01E20C01-01E

8-13

8. OPERATION VIA HART COMMUNICATOR

Example 2: For adjustments using a voltmeter

F0816-1.EPS

AXFA11 : FI-1201
Adjustment
1 Auto Zero Exe
2 Magflow Zero
3 D/A trim
4 Scaled D/A trim

HELP

SAVE

HOME

ENTER

1

2

Select the “Scaled D/A trim”.

Press OK (F4).

AXFA11 : FI-1201
WARN-Loop should be
removed from
automatic control

DEL

DEL

ABORT

OK

F4

JKL

4

(OK)

F0816-2.EPS

AXFA11 : FI-1201
Setting fld dev
output to 4mA

HELP

SAVE

ABORT

OK

8

VWX

2

AXFA11 : FI-1201
Trim will be scaled
from 4.000 to 20.000
1 Proceed
2 Change

HELP

SAVE

ABORT

ENTER

3

4

Select “Change”.
The same operations as for “D/A trim” are required when
selecting “Proceed”.

Enter the reading value on the meter when the signal is 4 mA.
In this case, enter the value of the voltage across a 250

resistor (1 V), and press ENTER (F4).

Enter the reading value on the meter when the signal is 20 mA.
Then, enter “5” (5 V), and press ENTER (F4).

Select “Proceed” and press ENTER (F4).

Connect the voltmeter, and press OK (F4).

Press OK (F4). The output signal of 0% is output.

AXFA11 : FI-1201
Set scale- Hi output
value
20.000000
20.000000

HELP

DEL

ABORT

ENTER

5

AXFA11 : FI-1201
Set scale- Lo output
value
4.000000
4.000000

HELP

DEL

ABORT

ENTER

AXFA11 : FI-1201
Trim will be scaled
from 1.000 to 5.000
1 Proceed
2 Change

DEL

DEL

ABORT

ENTER

6

7

AXFA11 : FI-1201
Connect reference
meter

HELP

DEL

ABORT

OK

‘1’

F4

(ENTER)

‘5’

F4

(ENTER)

F4

(ENTER)

F4

(OK)

F4

(OK)

STU

1

or

Advertising