Pressure test

Oxygen pressure

Information screen: oxygen pressure is too low.

NOTE:

The calorimeter C1 required for functionality an oxygen

pressure of at least 20 bar.

Reproducible measurements for a working pressure

of 30 bar is required.

System check

System start when the device is closed.

The system is brought through the system check in a safe

and completely deflated state. For this, the oxygen and water

discharge is performed.

Here, the following information screen is displayed:

INFO: During the system check detected no oxygen or the oxygen

pressure is too low, the following error message is displayed:

A pressure test must be performed after a decomposition vessel

has reached or exceeded the recommended number of ignition

cycles.

Once the pressure test has been performed a release code can be

entered to enable the unit to be used for further measurements.

The warning message will then disappear. The enable code can

be in the following screen, by pressing the „Next button“ to be

entered.

INFO: Using the unit may be further worked.

Contact for performing the pressure test, the

service department from IKA Works.

Observe the safety instructions!

[email protected]