Technical data, Maintenance warranty – IKA C 1 Package 1/10 User Manual

Page 40

Advertising
background image

40

Technical data

Power supply unit (external)

Design voltage /-frequency

100 - 240 V AC 50/60 Hz

Power consumption max.

120 W

Calorimeter:

Design voltage

24 V DC 5A

Power consumption max.

120 W

Permissible duration of operation

100 %

Protection tzype acc. DIN EN 60529

IP 20

Protection class

I

Overvoltage category

2

Contamination level

II

Permissible ambient temperature

5 - 40 °C

Permissible relative humidity

80 %

Operation at a terrestrial altitude

max. 2000 m

Dimensions (B x T x H)

290 x 300 x 280mm

Weiht

15 kg

Measuring range (max.)

40.000 J

Mode of the measuring

isoperibol 22 °C

isoperibol 30 °C

Measuring time

approx 7 min

Reproducibility (1g benzoic acid NBS39i)

0,15 %RSD

Working temperature

20 - 30 °C

Temperature measurement resolution

0,0001 K

Cooling medium temperature

19,5 K +/- 1,5 K for measuring mode isoperibol 22 °C

27,5 K +/- 1,5 K for measuring mode isoperibol 30 °C

Cooling medium permissible operating pressure

1,5 bar

Cooling medium

tap water / drinking water quality

Cool art

flow rate

Flow volume

50 ... 60 l/h

Oxygen operating pressure

30 ... 40bar

Interfaces

RS232

USB-device

Subject to technical changes!

Maintenance

Warranty


The device is subject only to the natural wear and tear of compo-
nents and their statistical failure rate. According to the IKA ma-
nufacturer specifications we recommend an annual maintenance
with security check and pressure test according to Pressure Equip-
ment Directive!
Cleaning

For cleaning, disconnect the main plug!

Only use cleaning agents recommended by IKA

®

.

Contamination Cleaning agents
Dyes

isopropyl alcohol

Building materials

water containing tenside/isopropyl alcohol

Cosmetics

water containing tenside/isopropyl alcohol

Foodstuffs

water containing tenside

Fuels

water containing tenside


- When cleaning, please do not allow moisture to get into the unit.
- Wear the proper protective gloves during cleaning of the devices.

- Before performing a non-recommanded method of cleaning or

decontamination, the user must ascertain with IKA

®

that this me-

thod does not destroy the instrument.

Spare parts order
When ordering spare parts, please give:
- Machine type
- Serial number, see type plate
- Software Version (second displayed value when the device is on)
- Item and designation of the spare part please see www.ika.com.
Repair
Please send instrument in for repair only after it has been

cleaned and is free from any materials which may consti-

tute a health hazard.

For this you should request the “Decontamination Clearance

Certificate” from IKA

®

, or use the download printout of it from

the IKA

®

website www.ika.com.

Return the instrument in its original packaging. Storage packaging
is not sufficient. Also, please use suitable shipping package materi-
als.

In accordance with IKA

®

warranty conditions, the warranty period is 12 months. For claims under the warranty please contact your local

dealer. You may also send the machine directly to our factory, enclosing the delivery invoice and giving reasons for the claim. You will
be liable for freight costs.
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insufficient care or maintenance not
carried out in accordance with the instructions in this operating manual.

C1 092014

Advertising
This manual is related to the following products:

C 1