IKA C 1 Package 1/10 User Manual

Page 23

Advertising
background image

23

Phase 9:

The system is completely emptied, and can be opened.

The inner vessel can be taken down. Through a visual check

is necessary to ensure that the sample in the crucible was

combusted completely.

After each combustion are all parts of the inner vessel that come

in contact with the combustion products to clean condensate.

It is sufficient to parts with an absorbent, lint-free cloth to wipe

off. Combustion residues in the inner vessel, for example, soot

or ash are eliminated in the same way. Can not be cleaned in

the manner described (eg due to burn-in or pitting corrosion) of

Cleaning the inner vessel

the inner vessel, it shall be subjected to no cases of mechanical

cleaning.

Please contact in such cases, contact your local service center or

send the inner vessel for inspection and cleaning to our factory.

Follow-up measurement

By pressing the multi-function element, the result display returns

to the measurement sub-menu and the next attempt can be

started.

Use of a disposable crucible

The disposable crucible can optionally be used instead of metal

or quartz crucible and burned during the measurement complete

(see chap. „Accessories“).

INFO: 50Joule omitted. With the combustible crucible not a

cotton thread is used.

Display the storage overview

If after pressing the „New Measurement“ button, the memory

overview is displayed, the capacity of the storage spaces for

measurements / calibrations are almost depleted. If the maximum

number is reached, no new measurements / calibrations more can

be saved for this mode. By pressing the OK key to access the screen

„New Measurement“. Delete measurements / calibrations are no

longer needed.

Result Display

NOTE:

Because the calorimeter C1 (all internal parts) still not in a

first measurement at operating temperature, the result of

the first measurement can differ significantly from the

subsequent measurements of a test series.

We therefore recommend not to evaluate the outcome and

possibly to delete.

Alternatively, a system test can be performed,

see chap. „System Test“.

C1 092014

Advertising
This manual is related to the following products:

C 1