System test, Measurement archiv – IKA C 1 Package 1/10 User Manual
Page 26

26
System test
Functional testing of the system
Use for the system test one or two benzoic acid tablets.
Alternatively, however, for example, an appropriate amount of
sugar or an alternative fuel to be used.
The system test is used to test the functionality of the overall
system. For this purpose all switching operations are carried out,
which take place in a normal measurement. If the system test
performed without error, the functionality is guaranteed.
The system test can be used to create a quick „warm up“ the
system. It is faster than a normal measurement, since there is no
analysis and improves the accuracy of the first measurement so
that they need not be discarded
Measurement archiv
In measurements Archive up to 100 measurements can be stored.
A measurement taken is automatically added to the archive.
Displays the line number, the name of the measurement as well
as the result.
Further details are displayed by clicking on a measurement.
C1 092014