Change seal:

1 In the Menu Maintenance press „change seal“ button. This

requires oxygen pressure be present at C1.

2 Unscrew the piston (Pos. 38 III) with screwdriver

counterclockwise

3 Remove the O-ring (Pos. 38 VI)

4 Put new O-ring on piston (Pos. 38 VI)

5 Screw in piston and tighten securely

6 „Change seal“ button again actuate pistons move to the

home position

38

38

III

Stirrer test

The stirrer is switched on and off for a visual check.

O

2

System blow:

1 Lift out the inner vessel from the bottom and set aside

2 After about 20 seconds the button „O2 system blow“ button.

Keep openings with a paper towel or similar covered by hand

3 After about 3 seconds, this state switches itself off again.

This any remaining condensed water and dirt is

in the line blown out.

This screen shows:

the software and firmware version

the serial number of the device

the date of the last service

the date of the last pressure test

the number of performed ignitions

the number of ignitions in the next pressure test must be

performed

the number of performed ignitions and support information

IKA - Contact information

blow

O-ring

(Water inlet valve

seal)

Informations

Stirrer

Servo up

The locking pin is extended.

Servo down

The locking pin is retracted.

VI (

8,0x2,5)

C1 092014

locking pin