Power supply unit connection

1. Connect the power supply
connecting cable (24V)

NOTE: Note the correct inser-

tion position (flat side

of the connector

pointing to the right).

Connecting cable

The desktop power supply (scope of delivery) must not be placed
flat on the laboratory table. It must be protected against moisture
and water penetration and must not be wet.

2. Tape the supplied rubber feet
to the power supply.
Place the power supply on the
rubber feet.
Connect the connection cable
to the DC power supply to
your supply network
(100-240V AC 50/60Hz).

flat side of the connector

Connection of peripherals

Heating:

Control output for the connection of the IKA

®

heating.

Scale:

RS232 interface for connection of a scale

(Mettler, Ohaus, Sortorius, Kern)

Interface parameters:

Baud rate: 1200

Data its:

7

Parity:

odd

Stop its:

1

Handshake: none

PC/Printer: RS232 interface for connecting a PC to control

the C1 (CalWin C6040) or a printer to output the

measurement data.

Interface parameters:

Baud rate: 9600

Data its:

8

Parity:

none

Stop its:

1

Handshake: none

USB-B:

USB device interface for connection the calori-

meters to the PC (CalWin C6040).

The connection simulates a serial interface on the

PC.

Installation:

After the C1 was connected with the supplied

data cable to the PC, the C1 tells the Windows

operating system, which device driver is required.

24V =

Programming (Prog)

interface

USB-B (USB)

PC / Printer (PC)

Scale (balance)

Heating (ext)

This driver can be downloaded from the

IKA website.

Find your driver at the following website:


http://www.ika.com/ika/lws/download/stmcdc.inf

Programming- (only for service)

interface:

24V=

24V Voltage input for the included power

supply table.

Handling the unit

1. Opening the unit

The closing of the unit in the reverse order

Q

h

h

1

2

Q

C1 092014