Calibration archive: C-Value selection

Click on the individual calibrations they are selected and marked

with a green tick.

By clicking again the selection is canceled and the new C-value

(mean value) is displayed below the list view directly.

By clicking the „OK“ button, the following user message appears:

By clicking the „OK“ button, the new value is accepted.

INFO: For RSD calculation of at least two C-values must be

selected. Otherwise, the result is Infinite and it is „Inf“ is

displayed as RSD.

Formula for calculating: X is a set of data (x1, .... xn)

and N the number

Relative Standard Deviation (RSD)

= (standard deviation / mean value) x 100

Mean value:

X = X

Mean

= (Σ

N

n=1

x

n

)/N

Standard deviation:

σ =

Σ

N

n=1

(x

n

- X

Mean

)

2

/(N - 1)

INFO: The new value is also displayed under Main menu "

Basic settings.

Calibration archive: Print

Click on the individual measurements they are selected and

marked with a green tick. By clicking again the selection is

reversed.

By clicking on the OK button, the following user dialog appears:

By confirming with „OK“, the selected calibrations are printed.

Calibration archive: Delete

Click on the individual measurements they are selected and

marked with a green tick. By clicking again the selection is

reversed.

INFO: Calibrations that are used to the current C-value

calculation, can not be deleted.

She are grayed out and can not be selected..

By clicking the „OK“ button, the following user dialog appears:

By confirming with „OK“, the selected calibrations will be deleted.