Selection of accessories

• Heating:

Use of heating during the measurement C1.2

• Combustible crucible:

yes / no

Weight input of the combustible crucible in the submenu „New

measurement“ can be done manually or by data transfer from

the balance.

• Printer:

Using a serial printer.

Selection option between:

• off: no expression

• briefly: expression

• standard: result expression with temperature values

• Service Information:

Service information is outputted via the serial interface

• Balance:

Selecting a balance type.

Selection option between:

out

Sartorius/Kern

Mettler-Toledo

Ohaus

Accessories

Chiller test

Maintenance

Viewing maintenance programs that can be performed by the

user at regular intervals.

Inner vessel and the inner vessel bottom lift out and set

aside.

Maintenance menu:

The system must be open to performing the maintenance programs.

The maintenance menu can be reached at Menu > Maintenance and contains the following points

Check whether the cooling water tem-

perature within the specified range of

19,5 +/- 1,5 °C at the wor-

king temperature 22°C

or

27,5 +/- 1,5 °C at the wor-

king temperature 30°C

is located.

C1 092014