After input a new measurement

User information: Memory

Prior to input a new measurement

Information about the available space when you select the menu

item „New Measurement“

This information is displayed when there is a value in the

critical region is located, that is, from 95 out of 100 possible

measurements in memory or from 13 of 15 calibrations in store

for IB1, IB2 at 22 ° C or 30 ° C operating temperature.

After entering the values for a new measurement archive

database will be checked again and if no free space is available

following message is displayed:

Depending on the selection measurement or calibration, the

screen sample will vary.

The user can skip this message, then the following measurement or

calibration is not saved. However, it can also switch on the button

„Archive“ to the appropriate archive and then delete old or unused

measurements or calibrations.

Exchanging of wear parts

O

2

-Valve seal change

Use only original IKA spare parts!

1 Bushing (Pos. 33) with the enclosed tool SW11

screw out

2 Lift piston (Pos. 32/21) with tweezers

3 Remove the O-ring (Pos. 40), insert a new O-ring and ensure

the correct position

4 Set piston (Pos. 32/21) back in the opening

5 Bushing (Pos. 33) screw with the enclosed tool SW11

handtight, paying attention to correct position of the

O-ring (Pos. 32/8)

33

32/8

32/21

40

(10,0x2,5)

(3,0x1,5)