IKA C 1 Package 1/10 User Manual

Page 19

Advertising
background image

19

Input window NEW MEASUREMENT

Select and confirm the menu item „Sample weight“. Enter the

sample weight with numeric keypad and confirm.

INFO: A distinction is made between the use and non disposable

cup. Use of disposable crucible can be selected in Settings

" Accessories.

INFO: For automatic input of the weight may be a connected

scale can be used

INFO: The inputted weight values are checked for permissibility.

INFO: The existing store will be reviewed after the entry of a

new measurement / calibration and a user note displayed

if it is not possible to save the measurement / calibration.

NOTE:

Ensure that a maximum energy input of from 40,000 joules is not

exceeded.

NOTE:

For more, detailed descriptions of the individual menu items are

treated in the chapter „Calorimetric Measurements“.

INFO: If a scale is used in activation of the input field, the

weight is automatically queried. It can be pressed on the

scale alternatively, the transmit button.

During the prerequisite checking, the cotton thread can be

attached to the ignition wire, hung the crucible in the provided

holder and the sample are contacted with the ignition wire.

Placing the sample

When using a disposable crucible of contact between the ignition

wire and the combustible crucible must be present. The fixing of

the cotton thread is not necessary.

2. Weigh the substance with an accuracy of 0.1 mg directly into

the crucible.

Where appropriate distilled water or a solution must be

presented in the inner vessel.

The input of the maximum possible sample weight is limited to a

maximum of 5g.

1. Attach with a loop a cotton thread centered on the ignition

wire (at the open calorimeter).

The inner container base remains in the control of sample prepara-

tion in the calorimeter. But you can also use it by pulling it out

individually, or for consequential - Analysis - determinations, refer

to the closed inner container.

The following preparations must be made to prepare the system

for a measurement:

Preparing a measurement

C6 Quartz crucible

h

Impression for distilled water or a
solution as a template

C1 092014

Advertising
This manual is related to the following products:

C 1