E02 COMMUNICATION

If no internal communication between the software and firmware

are prepared COMMUNICATION ALARM appears..

E04 MINTEMPDIFF

If 30 seconds is no temperature increase of 0.5 ° C takes place

after ignition, the alarm is displayed.

The most common causes for this problem are:

the cotton thread is not in contact with the sample

the contact with the fuse wire is not present due to

contamination or an not properly tightened screw

the bad combustibility the sample, optionally you must use

firing aids

Low oxygen pressure. Check the oxygen supply (30-40 bar)

Short circuit at the ignition electrode, crucible holder is

present at the ignition electrode

E05 FILLWATER

Message appears when the system during filling with water after

50 seconds does not detect water in the upper water outlet.

Causes for this problem are:

the cooler is not switched on or not connected

Water valve seal defective

no oxygen present (for switching the valves, the system

requires pressure)

upper water sensor is defective

the filter in the supply line is dirty

Water valve tightly or defective

E03 DRIFT ALARM

Drift Alarm

In preliminary experiment or or post-experiment the temperature

drift criterion is not met.


Check the stirring function.