4.10 Oxygen station C 248

The C 200 does not have an integrated oxygen
fi lling point for the decomposition vessel. You
can use oxygen station C 248 for fi lling. The C
248 oxygen station must be set up at least 1.5 m
far away from the calorimeter. For information on
how to operate and connect the oxygen station,
please see the accompanying operating instruc-
tions.

Operation:

• Place decomposition vessel C 5010 on the

marked position.

• Move the lever  down and centre the de-

composition vessel below the fi ll head .

• Lock the lever in place.
• Fill the decomposition vessel for approx. 30

secs.

• Move the lever back to its original position.

4.8 Switching off the system

Standby mode
In order to switch off the calorimeter system, the
start screen must be displayed.
Press OFF (F1). The appliance will switch to

standby mode.

Switching off
Only switch the appliance off in standby mode.
To do this, press the ON/OFF switch  (see sec-
tion 4.2) on the back of the appliance.

4.9 Coding decomposition vessels

You can use several decomposition vessels (max.
4) when working with calorimeter C 200. You
can code these to distinguish between them by
sticking the black coding strips into the special
slots on the decomposition vessels.

TIP

Slot for

coding strips

1

2

3

4

Decomposition vessel no.:

without coding

Lever