10 technical data – IKA C 200 User Manual
Page 30

30
Table power supply (external):
Rated voltage
100 - 240 V AC
Frequency
50 / 60 Hz
Input power max.
150 W
Calorimeter:
Rated voltage
24 V DC 5A
Input power max.
150 W
Fuses (internal)
1x 2,5 AT
Perm. on-time
Continuous operation
Protection class as per DIN EN 60 529
IP 20
Protection class
III
Overvoltage category
2
Contamination level
II
Perm. ambient temperature
20 °C ... 25 °C (constant)
Perm. ambient humidity
80%
Operation at a terrestrial altitude
2000 m above sea level
Dimensions
400 x 400 x 400 (W x D x H)
Weight
21 kg
Measurement ranges
40.000 J
Measuring mode / measurement times
Isoperibol / approx. 17 min
Dynamic / approx. 8 min
Time control / 14 min
Manual / approx. 17 min
Interfaces
1 x parallel (Centronics)
1 x serial (RS 232)
Subject to technical changes!