7 service menu – IKA C 200 User Manual

Page 26

Advertising
background image

26

Please note that you must stop every started ac-
tion.

Exceptions:
The "Ignite” action ends automatically after a set
time period (approx. 2 seconds).
With actions "Reset” and "ReInit” you will au-
tomatically exit the menu after the actions have
been performed.

If you exit the service menu, all the started actions
will be stopped and the system will be restored to
its original state. This will allow you to continue
without any errors.

7.2 Description of service menu options

Ignition

This menu item allows you to check the ignition
function.

Requirement:
There must be a decomposition vessel without
sample, but with ignition thread in the inner ves-
sel and the cover must be closed.

Fill IV

This menu item allows you to fi ll the inner vessel
manually.

Requirement:
There must be enough water in the tank.

Empty IV

This menu item allows you to empty the inner
vessel.

Requirement: the emptying hose must be locked
in place in the plug-in coupling (section 4.2).

Stirrer

This menu item allows you to switch on the stir-
rer drive and check that the magnetic rod is also
rotating in the inner vessel.

Requirement:
There must be approx. 0.5 l water in the inner
vessel.

Meldung:
Füllzeit überschritten

7.1 Operation

This menu allows you to directly control and test
various actions and statuses of the calorimeter
without performing a measurement. There are
also menu items which can be used to start up
and shut down the appliance. The service menu
can only be run if the calorimeter displays the
start screen.

Perform the desired action using:

MENU (F3)

UP/DOWN (F2) to "Service"

OK (F1)

There are now eight actions to choose from

Select the desired action with UP/DOWN (F2)

OK (F1) starts the action

OK (F1) stops the action again

7 Service menu

Advertising