IKA C 200 User Manual

Page 28

Advertising
background image

28

8.4 Maintaining the water circulation

In the case of discontinuous use with tap water
(single measurements with long gaps) a stabiliser
must be added to the water circulation to prevent
the formation of algae.
Add approx. 4 ml IKA

®

-Aqua-Pro C 5003.1 to the

tank (see Accessories).
Switch the pump on via the service menu (sec-
tion “Pump“). Switch the pump off again after
30 - 60 s.

You can also add the 4 ml Aqua-Pro to the water
maintained at a constant temperature in the last
measurement (before long gaps in measuring).

If the appliance is not going to be in opera-

tion for a long period of time, it is advisable

to completely empty the calorimeter's water

circulation.

The water must be drained out before trans-

portation.

Empty the outer vessel via the service menu (sec-
tion 7.2 “Outer vessel“). The emptying hose must
be locked in place in the plug-in coupling  (sec-
tion 4.2).
Empty the tank by locking the emptying hose in
place in the plug-in coupling  (section 4.2). The
tank will empty automatically.
Press the locking knob on the plug-in coupling 
to remove the hose.

8.5 Decomposition vessels

Please see the operating instructions for

C 5010 for information on decomposition

vessel maintenance!

8.6 Cleaning information

Only clean IKA

®

appliances using these IKA

®

ap-

proved cleaning agents:

Dirt

Cleaning agent

Dyes Isopropanol
Building materials

Water containing

detergent,
Isopropanol
Cosmetics

Water containing

detergent,
Isopropanol
Food

Water containing

detergent,
Fuels

Water containing

detergent,
Other materials

Please consult IKA

®

• For cleaning the micro fi lter you must empty

the outer vessel (see section 7.2).

• Screw the cover of the micro fi lter down and

pull you out the fi lter element.

• Clean the fi lter element with clear water.

• Insert the fi lter element after cleaning with the

opening inward again into the housing of the
micro fi lter and screw you on by hand the co-
ver.

• Fill the outer vessel again with water (see sec-

tion 4.7).

• Control the tightness after you the equipment

again fi lled (see section 6.6 “Fill time exceeded).

TIP