8 cleaning and maintenance – IKA C 200 User Manual

Page 27

Advertising
background image

27

Reset

This menu item allows you to restore the default
settings.

Reinit

This menu item allows you to reinitialise the
A / D converter.

8.3 Micro fi lter

In the back plate of the equipment is additionally
a micro fi lter. If the micro fi lter is dirty, the water
fl ow in the equipment is reduced. Thus the water
fi lling time of the inner vessel extends. Thereby
if the upper time limiting is exceeded by 120 se-
conds, the system indicates the disturbance and
breaks the measurement off.

Pump

Switch the pump on with this menu item. The
outer vessel will be fi lled and rinsed (see also sec-
tion 4.7).

Requirement:
Check that there is water in the tank.

Empty outer vessel

This menu item allows you to empty the outer
vessel.

Note:
Shut-down, completely empty appliance.

Requirement:
The emptying hose must be locked in place in
plug-in coupling  (section 4.2).

In order to ensure trouble-free operation over a
long period of time, it is important to carry out
the following maintenance work to the calorime-
ter system.

8.1 Inner vessel fi ller

Remove the fi lter element from the inner vessel
and clean the inner vessel fi lter in clear water or
an ultrasonic bath if there is any dirt visible. Also
clean the inner vessel, by re-moving the positio-
ning mount for the decomposition vessel. Place
the fi lter back on the inner vessel connector after
cleaning.

Operation without the fi lter element causes dirt
in the valves and thus appliance failure.

8.2 Filler

There is also a fi lter in the fi ller. This prevents dirt
in the storage container. If the fi lter is dirty, remo-
ve the entire fi ller and clean the fi lter (see section
4.1, ).

8 Cleaning and maintenance

Advertising