IKA C 200 User Manual

Page 20

Advertising
background image

20

Fill the decomposition vessel using oxygen sta-

tion C 248 (see section 4.10).

Slide the ignition adaptor onto the decompo-

sition vessel.

For information on how to operate the decompo-
sition vessel, please see the operating instructions
provided.

6.2 Preparing the decomposition vessel

Prepare the decomposition vessel as follows:

Unscrew the union nut and remove the cover

using the handle.

Attach a cotton thread to the centre of the

ignition wire using a loop.

Weigh out the substance directly into the

crucible with an accuracy of 0.1 mg. Note the
weight or enter directly into the calorimeter. (See
section 6.3 "Preparing the measurement").

Insert the crucible into the crucible holder.

If necessary put some distilled water or a solution
into the decomposition vessel.

See section 5.3 "Information about the samp-

le" and section 1 "Safety precautions”.

Using tweezers, align the cotton thread so that

it hangs inside the crucible and is immersed in the
sample. This will ensure that the burning thread
ignites the sample during the ignition process.

Attention! Place the cover onto the lower

section and push down until it presses against the
stop piece in the lower section. Place the union
nut onto the lower section and tighten by hand.

Cover

Union nut

Handle

Cotton thread

Max. 1 mm