Installing Electrically

The T-Guard requires a 24 V DC power supply (tolerance ± 10 %).
A wrong polarity triggers the fuse.
The power supply should be able to supply a current of maximum 6 A. Typically, the
T-Guard needs 70W.

The following maximum cable lengths have to be met:

Notice

Do not power up the power supply before having attached all other cables
you need!

Power Connector

Only two of the four pins are used for the power supply. The plus pole is marked with
„1+“, the minus pole is marked with „2-“.

4.5

Connecting the control unit

Attach one end of the control unit cable to the connector CONTROL UNIT of the T-
Guard™ Leak Detection Sensor and the other end to the control unit itself. Now T-
Guard™ Leak Detection Sensor can be operated via the control unit.

0.75 mm²

8.5 m

AWG 18

10 m

1 mm²

10 m

AWG 16

15 m

1.5 mm² 17.5 m

AWG 14

25 m

Caution

The CONTROL UNIT output of the T-Guard™ can be damaged when the cable is
disconnected or connected during operation.

Only plug or pull the cables when the T-Guard™ is switched off.