2 explanation of menu items, Explanation of menu items – INFICON T-Guard Leak Detection Sensor User Manual

Page 58

Advertising
background image

56

Operation

jina

85

e 5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

11

2)

5.8.2

Explanation of Menu Items

The menu items the explanation refers to are written in bold letters.

Main Menu

Back

Returns to the previous page, does not alter any setting.

Main menu

Display

Contrast

Here, you can change the image contrast of the display.

Possible values: 0 to 99.
Default setting: 25.

Contrast

Invert display

Inverts dark characters and light background to light characters and dark
background.

Lower Limit

Raises the lowest leak rate displayed.
Increase by factor 1 to 100 possible. Default setting: 2

Main menu

Operating mode

Accumulation

Chooses Accumulation as operation mode after confirming it with OK.

Carrier gas

Chooses Carrier Gas mode for operation after confirming it with OK.

Continuous Mode

Chooses Continuous mode for operation after confirming it with OK.

Standby

Elongates sensor life time. Press START to restart.

Main menu

Trigger

Trigger level 1

Enter the reject leak rate here.

This setting determines the measuring time and decides between gross and fine
measurement.

Possible values: 5E-6 mbar l/s to 9.9E-2 mbar l/s.
Default setting: 2E-5 mbar l/s.

Trigger level 2

Advertising