4 electrical connection of the plc outputs, Electrical connection of the plc outputs, Caution – INFICON T-Guard Leak Detection Sensor User Manual

Page 39

Advertising
background image

Installation

37

jina

85

e 4

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

11

2)

4.6.4

Electrical connection of the PLC outputs

The default setting is as follows:

Pin

Command

Explanation

8

Trigger 1

9

Error/Warning

10

Ready

11

Measure

12

Standby

13

Rec. Strobe

14

Warning

15

Cal active

16

Common

Common connection for all PLC outputs.
This contact can be used as the positive or the negative
pole.

Caution

The maximum permissible voltage for the output signals is 28 V.
The maximum total current for the output pins 8 - 11 is 0.75 A.
The maximum total current for the output pins 12 - 15 is also 0,75 A.

Permissible output current

Max. 125 mA per output

Load resistance of outputs

Min. 130 ohm per output

Supply voltage of
outputs

24 V DC

Fig. 5-4 Electrical connection of the PLC outputs

16

8

9

10

11

12

13

24V

14

15

PLC IN/OUT

Advertising