4.6.4
Electrical connection of the PLC outputs
The default setting is as follows:
Pin
Command
Explanation
8
Trigger 1
9
Error/Warning
10
Ready
11
Measure
12
Standby
13
Rec. Strobe
14
Warning
15
Cal active
16
Common
Common connection for all PLC outputs.
This contact can be used as the positive or the negative
pole.
Caution
The maximum permissible voltage for the output signals is 28 V.
The maximum total current for the output pins 8 - 11 is 0.75 A.
The maximum total current for the output pins 12 - 15 is also 0,75 A.
Permissible output current
Max. 125 mA per output
Load resistance of outputs
Min. 130 ohm per output
Supply voltage of
outputs
24 V DC
Fig. 5-4 Electrical connection of the PLC outputs