74

Errors and Warnings

jina

85

e 9

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

11

2)

9

Errors and Warnings

Error No.

Plain text message

Possible reason

Recommendation

E1

24 V of the MC50 is too low

Fuse F1 on the DC/DC board has
blown or power supply unit too
weak

Replace the Fuse F1 on the DC/
DC board or use a controlled
power supply unit.

E2

24 V of sensor heating too low

Fuse F2 on the DC/DC board has
blown or power supply unit too
weak

Replace the Fuse F2 on the DC/
DC board or use a controlled
power supply unit.

E3

24 V II too low

Fuse F3 on the DC/DC board has
blown or power supply unit too
weak

Replace the Fuse F3 on the DC/
DC board or use a controlled
power supply unit.

W4

24 V at socket OPTION is too low

Fuse F4 on the DC/DC board has
blown or power supply unit too
weak

Replace the Fuse F4 on the DC/
DC board or use a controlled
power supply unit.

E7

-15 V of the MC50 is too low

DC/DC board defective

Call nearest INFICON service
representative

E8

15 V of the MC50 is too low

DC/DC board defective

Call nearest INFICON service
representative

E9

High Voltage error

Problem in sensor electronics

Call nearest INFICON service
representative

W11

Wise-current unstable

T-Guard™ Leak Detection Sensor
possibly not being used for
several days, sensor current
should stabilize after longer run-
time.
The helium background is higher
than 20 ppm.

Get the T-Guard™ run for 2
hours. Reduce the helium
background in the room.

E12

Wise Sensor not ignited

Current from Wise Technology
sensor too low for more than 10
min after power on.
(>5E-11 A).

Restart the T-Guard™ Leak
Detection Sensor, if problem
persists, call nearest INFICON
service representative

E13

Problem in Heater Control

Heater control for Wise
Technology sensor defective.

Restart the T-Guard™ Leak
Detection Sensor, if problem
persists, call nearest INFICON
service representative

E14

Discharge gone out

Sensitivity problem of Wise
Technology sensor
or
No ambient helium present (e.g.
inlets purged with nitrogen)

Restart with inlets connected to
external air

If problem persists, call nearest
INFICON service representative

E19

No communication with Wise ADC

Wise Technology sensor defective
or
CPU-board defective

Call nearest INFICON service
representative!

E20

Temperature at electronic unit is too
high (>60°)

Ambient temperature too high

Cool down environment, place
the T-Guard™ Leak Detection
Sensor in cooler area.

E20

Temperature at electronic unit is too
high (>60°)

Ventilation failure

Check if fan is (check for air flow
through inlet of the housing)

Advertising