INFICON T-Guard Leak Detection Sensor User Manual

Page 78

Advertising
background image

76

Errors and Warnings

jina

85

e 9

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

11

2)

W61

EEPROM parameter initialized!

Software update performed and
new parameters have been
introduced

Newly introduced parameters are
listed below warning

Acknowledge warning

If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty

Replace EEPROM*

W62

Lost EEPROM parameters!

Parameter has been modified
during software update and reset
to default value

Affected parameters are listed
below warning

Check setting of modified
parameters in corresponding
software menu and set to desired
value

If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty

Replace EEPROM*

W64

There are outstanding warnings!

Acknowledged but still valid
warnings will be repeated every 2
hours or on each new power on

Please double-check the
warnings

W81

Calibration factor too low!

If the calibration factor was
measured with 0.2, calibration
was not successful. If the
calibration factor was confirmed
between 0.2 and 0.5, a warning
will be output, however, also after
the confirmation of the successful
calibration.

The leak rate of the test leaks is
higher than entered.

Check correct setting of test leak
value

If the calibration factor was
measured to be higher than 5,
calibration was not successful. If
the calibration factor was
confirmed between 2 and 5, a
warning will be output, however,
also after the confirmation of the
successful calibration.

Check the settings of the
chamber volume or the gas flow.

W82

Calibration factor too high!

Calibration factor during
calibration determined to be > 10

The leak rate of the test leaks is
smaller than entered.

Check correct setting of test leak
value

Chamber volume or gas flow are
greater than entered.

Check the settings of the
chamber volume or the gas flow.

W84

Test leak signal too small

Test leak faulty

Please check all your settings.

Trigger value(s) too high

Check the trigger

Settings of volume/flow too low

Check the free volume/the carrier
gas flow

Test chamber leaky

Check the chamber for tightness

Error No.

Plain text message

Possible reason

Recommendation