4.6.3

Selection of functions assigned to the PLC outputs.

Main menu

Settings

Interfaces

Define PLC outputs

In the sub-menu D

EFINE

PLC

OUTPUTS

, you can set which pin

of the I/O connector (suitable for PLC outputs) shall be assigned to which command.

Signal

Explanation

OPEN

low-resistance, for testing purposes

CLOSE:

high-resistance, for testing purposes

TRIGGER 1

low-resistance means Trigger 1 was exceeded

TRIGGER 2

low-resistance means Trigger 2 was exceeded

READY

low-resistance means the T-Guard™ is ready for the next
measurement. In Continuous Measurement mode, the T-
Guard measures the reference line.

CYCLE ACTIVE

low-resistance means the T-Guard™ Leak Detection Sensor is
just performing a measurement cycle This includes automatic
purging and reference measurement in Carrier Gas mode.
There is no measuring cycle in Continuous Measurement
mode. That is the reason why the output is always high-resis-
tive.

STANDBY

low-resistance means that the T-Guard™ Leak Detection
Sensor is just in Standby mode

ERROR

low-resistance means an error has occurred

WARNING

low resistance means there is an outstanding warning

ERROR / WARN

low-resistance means an error has occurred or there is an
outstanding warning

CAL ACTIVE

low-resistance means a calibration is running right now

REC. STROBE

low-resistance means the recorder output is invalid for 300 ms
because of changing the decade of LR Mantissa and LR
Exponent.

MEASURE

low-resistance means that the T-Guard™ Leak Detection
Sensor is just measuring. The leak rate of the last measure-
ment is on when this output and also the REC STROBE output
are high-resistive.

GROSS / FINE

low-resistance means that the T-Guard™ Leak Detection
Sensor is just in GROSS mode

GROSSLEAK

low-resistance means the highest trigger value was exceeded
by at least the factor 5.

CONTAMINATED

low-resistance means the sensor was contaminated with too
much helium. In this case, immediately purge the reference
line with external air or nitrogen to decontaminate the sensor.

INV TRIGGER 1

high-resistance means Trigger1 was exceeded

INV TRIGGER 2

high resistance means Trigger2 was exceeded

INV READY

high-resistance means the T-Guard™ is ready for the next
measurement. In Continuous Measurement mode, the T-
Guard measures the reference line.