Errors and Warnings

75

jina

85

e 9

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

11

2)

W28

Real time clock reset! Please enter
date and time!

CPU-board has been replaced

Please enter date and time

Battery on CPU-board faulty

Replace CPU-board*

E32

Wise current too high

Wise Technology sensor current
exceeds threshold value

Do not switch off the T-
Guard™ Leak Detection
Sensor
Supply external air through the
sniffer line quickly, restart the T-
Guard™ Leak Detection Sensor
and wait to allow the T-Guard™
Leak Detection Sensor to
recover!
If problem persists, call nearest
INFICON service representative.

E38

Pump defect

Sensor pressure higher than 800
mbar
Fore pump defect, not switched on
or pump hose defect

Please check the fore pump.

W39

Valve block error!

Cannot recognize valve block
version during self-test

Call nearest INFICON service
representative

W41

Pressure difference too high

Pressure difference between
measuring inlet /air inlet ±10%
Pressure difference between
READY and FINE

Check inlet filters

W43

Pressure too low

Pressure below the lower limit.
Reference line plugged

Check the lower pressure limit
Check the reference line

W44

Pressure too high

Pressure above the upper limit
Wrong fore pump
Pump hose defect

Check the upper pressure limit
Please check the fore pump and
the pump hose

W45

Too high helium concentration

During the purging process, too
much helium is before or in the
measuring line. T-Guard stops the
purging process too early to
protect the sensor
The test chamber was not purged
sufficiently before the
measurement.

Purge the chamber with external
air and start a new purging
process of the sensor via the PLC
or the RS232 command,
Purge the test chamber more
thoroughly before
measurements.

W50

No I·Stick present.

I·Stick was not connected during
running-up.
I·Stick defect

Switch the T-Guard off and on
again with the I•Stick connected.

W52

Lost I·Stick parameters!
Please check the settings!

I·Stick disconnected during
operation
I·Stick defect

Please check all your settings.
Switch the T-Guard off and on
again with the I•Stick connected.

W59

EEPROM parameter queue
overflow!

May occur when the software has
been updated to an older version.

Restart the T-Guard™ Leak
Detection Sensor, if problem
persists, call nearest INFICON
service representative

W60

All EEPROM parameters are lost!
Please check your settings!

New EEPROM has been installed,
EEPROM on motherboard is
virgin

All settings in software menu are
reset to default! Please enter
your settings again

If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty

Replace EEPROM*

Error No.

Plain text message

Possible reason

Recommendation