Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions


OP-Sensistor ISH2000-EN-201203

23

EN

background concentration will be detected. However, if the concentration remains
constant it will be gradually cancelled out over a period of several minutes.

For example, if the background concentration for some reason should suddenly rise
to 10 ppm H

2

, then the detector will give a corresponding signal which will very

slowly decline to zero. If you thereafter expose the probe to a leak which gives rise
to another 10 ppm H

2

, then the detector will give essentially the same signal as if

there were no background concentration.

Sensitivity

Sensitivity

of audio signal and signal bar in

Detection Mode

.

Note:

This does not affect the

Analysis Mode

.

Auto Range

Set this parameter to

ON

for auto ranging of sensitivity in

Detection Mode

. Sensitivity

will decrease two steps if the signal reaches full scale. Sensitivity is restored to
selected

Sensitivity

(See “To locate leaks” on page 22) when signal returns to zero.

Direct Sensitivity Adjustment

Setting this parameter to

OFF

will remove the sensitivity adjustment from the

Detection Mode

display. Sensitivity can still be adjusted in the

Sensitivity Settings

menu

after entering password (if set).

Note:

The sensitivity setting only affects the

Detection Mode

.

Audio Threshold

Makes it possible to lower the sound to a set level in Detection Mode. The level is in
% of full Detection bar.

Reject Indicator

Makes it possible to show (not shown) the indication Reject in Detection Mode.

Audio Ready Pulse

This sets the standby sound to a silent or or pulsating tone.

8.6 Analysis mode

In

Analysis Mode

the measured value is displayed in figures. The default unit is in PPM

but it is possible to choose other units, See “Default parameters” on page 40.

To analyse leaks

The Analysis mode is used for measuring the size of a leak (or the concentration of a
gas sample). To be able to do this measurement and obtain correct values, the
instrument must first be calibrated using the calibration function.

In the Analysis mode the detector determines the gas concentration from the
change, as the probe goes from being exposed to background to being exposed to a
certain gas concentration. The detector does not continuously monitor the gas
concentration but takes just one reading instead. Another suitable alternative name
for this mode could be Sampling Mode. It is important to keep this in mind when
using the detector in this mode.