Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions

28

OP-Sensistor ISH2000-EN-201203

EN

Audio Base Frequency

This sets the lowest audio base frequency tone in Analysis and Detection Mode.

Set Clock

Real time set as hh:mm:ss. Hours and minutes can be adjusted. Seconds will
automatically be set to 00 when hours and minutes have been set. Clock runs even
when detector is disconnected from the power supply.

Set Date

Real Time Clock date set as YY-MM-DD. Clock runs even when detector is
disconnected from the power supply.

Printer Port

The Sensistor ISH2000 is equipped with a serial (RS232) printer port. See “Printer
port” on page 33.

Info

Contains information about software versions, Serial number, and Internet contact
information.

8.10 Service Settings

The Service Mode is reached by starting the instrument and at the same time hold
the right button down on the panel. After start a new main menu called Service
Settings will appear.

Show Password

Shows the chosen password in case the customer has forgotten the password.
Contact INFICON AB to have the code sent to you. See the web address under the
section-Info.

Probe System Reset

Reset all parameters into the probe to default settings. Contact INFICON AB to have
the code sent to you. See the web address under the section-Info.

System Reset

Resets all parameters to default settings. Contact INFICON AB to have the code sent
to you. See the web address under the section-Info.

Detector Signal Level

The

Detector Signal Level

is the level which the sensor is considered to have recovered

from the last gas signal. It decides when the

DET_SIGNAL output will come on. this

signal can be used to block the start of a calibration or new test cycle in semi and
fully automatic testers.

IF

DET_SIGNAL is high then this means that the sensor has detected hydrogen and

has not yet recovered.

Detector Signal Level

can be adjusted in the

Service Settings menu

. You can increase

Detector Signal Level

if you have many small disturbing signals. A high setting of

Detector Signal Level

gives better tolerance to “noise” gas signals at the expense of