Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions


OP-Sensistor ISH2000-EN-201203

39

EN

Measurements:

Panel cut out: 262 x 127 mm (5 x 10.3 inches).

Maximum panel thickness: 8 mm (0.3 inch).

Allow an extra 20 mm (0.8 inch) on left and right side for mounting brackets.

Depth of open space in panel is at least 15 cm (6 inches) for Sensistor ISH2000 to
fit.

Installation process

Table 8-13. Install the detector as follows

Fig 8-7. Fixing the mounting brackets.

See “Sensistor ISH2000 specifications” on page 43 for electrical connections.

Step

Action

1.

Cut hole in panel according to figure above, and
remove burrs.

2.

Check that rubber o-ring is in correct position in
the grove around the edge of the detector.

3.

Put the detector in place in the panel hole.

4.

Hold the panel in place while fixing the mounting
brackets to the detector (See Figure 8-7).

5.

Align the detector horizontally and lock it by
tightening the 4 set screws.

6.

Lock the set screws with the locking nuts.

7.

Connect ground terminal to protective ground of
cabinet.

Sensistor ISH2000 photo of brackets

Bracket to panel screw

Bracket to unit

Bracket