Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions


OP-Sensistor ISH2000-EN-201203

25

EN

It is possible to measure or search in (Combined Mode) for a leak withut registering
the measurement. The measurement value registers only when the staple moves
(Multipoint Analysis Time).

If a dynamic number of measurement points is selected the do the measurement in
the following teps:

1. Push the probe button to begin the first measurement.
2. Place the probe near the measurement point during the time the staples move

(Multipoint Analysis Time).

3. You might need to wait until the next measurement. The instrument signals-

Wait

.

4. Repeat the procedure for the next measuring point.
5. When you sum up all the measurements push and hold on the probe button a

short time.

Multipoint Analysis Time

Set the time for each measurement.

Min Presentation Time

Signal values in

Analysis Mode

will never be presented shorter than this time. Values

are, however, always presented until the signal has recovered. The default value is 1
second, but values from 0 - 120 seconds can be used.

Display Threshold

Hides all measurements under a set % of Reject Level.

Audio Threshold

Instrument is silent under a set % of Reject Level.

Reject Indications

There are three choices of Reject Level indications except the LEDs indication:

Flashing screen

Chopped audio signal

Combination of indication 1 and 2.

Show Reject Level

Shows the Reject Level value on the display.

Audio Ready Pulse

This sets the standby sound to a silent or or pulsating tone.

8.7 APC Settings

APC is an abbreviation for Active Probe Control. The APC function is for the control
of an active probe that has a built-in alarm, valves or pumps via Probe Control Port.