10 sensistor ish2000 specifications, Sensistor ish2000 specifications – INFICON Sensistor ISH2000P Hydrogen Leak Detector, Panel Model User Manual

Page 45

Advertising
background image

Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions


OP-Sensistor ISH2000-EN-201203

43

EN

10 Sensistor ISH2000 specifications

Table 10-1. Power supply specifications.

Table 10-2. Input and output connections.

Power

Sensistor
ISH2000

Sensistor
ISH2000C

Sensistor
ISH2000P

AC mains voltage

100-240 V 50/
60Hz.

100-240 V
50/60 Hz

-

AC mains current

Typically 1 A (2 A
pulse at power
on).

Typically 300 mA

-

Fuse

2 A slow/ 250
VAC.

-

Nominal battery
voltage

-

16.1 VDC

-

Operating time

-

h without screen
saver, at 20 C.

-

Charging time

-

6.5 h

-

Power supply
voltage

-

-

24 VDC

Power supply
current

-

-

3 A max.

Input/output

Sensistor
ISH2000

Sensistor
ISH2000C

Sensistor
ISH2000P

Power input
connector

AC input connector,
IES 320.

Charger input
connector, 2.1 x 5.5
mm std. Positive
centre.

4 pin Phoenix MC 1.5/
5.81 Series
Detachable screw
terminal.

Probe control/Status
port

25-pin D-sub female.

-

25-pin D-sub female.

Minimum pulse length

40 ms

-

40 ms

Input impedance

50k ohm

-

50k ohm

Input maximum range

-34 to +38 VDC

-

-34 to +38 VDC

Input high

> 12.0 VDC

-

> 12.0 VDC

Input low

< 8.0 VDC

-

< 8.0 VDC

Output current

max 0.5 A/output,
max 2.5 A total

-

max 0.5 A/output,
max 2.5 A total

Inductive loads

External clamp diodes
recommended

-

External clamp diodes
recommended

Low state voltage

Max 1.5 VDC

-

Max 1.5 VDC

Low state leakage
current

Max 100 µA

-

Max 100 µA

Short circuit
protection

Thermal and
electronic

-

Thermal and
electronic

Advertising