Rs232 serial communication – INFICON Sensistor ISH2000P Hydrogen Leak Detector, Panel Model User Manual

Page 38

Advertising
background image

Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions

36

OP-Sensistor ISH2000-EN-201203

EN

The data format for Detection data output contains of ten ASCII characters. Nine
characters show the value in engineering format (See “Engineering format” on
page 17), and one character shows linefeed (LF).

The print time is 50Hz continuous streaming data.

Note:

Data rate is 25Hz when Sensistor ISH2000 is running in

Service Display Mode

.

RS232 serial communication

The most commonly used Sensistor ISH2000 functions can be started/configured
over the RS232 interface.

Table 8-10. RS232 communication specifications.

RS232 interface commands

Table 8-11. Common used functions.

K = Calibration request

Starts calibration if the Sensistor ISH2000 has an active probe driver installed.
Sensistor ISH2000 answers with a "K" if an active driver containing a calibration
routine was found and "F" if the calibration APC sequence not found. Calibration
doesn’t start if Purge level is reached.

M = Measure Request

Parity

None

Flow control

None

Setting

Value

Setting

No Printer

PC Printer

Data
Output

Data rate

115200 baud

1200 baud

9600 baud

Data bits

8

8

8

Stop bits

1

1

1

Parity

None

None

None

Flow control

None

None

None

Command

Header

Calibrate

K

Measure M

Print Request

N

Stop Measurement

Q

Hand Probe

R

Active Probe (Installed AP)

S

Analysis Mode

X

Detection Mode

Z

Combined Mode

Y