Probe Control Port Connector
The control port connector is a 25-pin female D-sub. Refer to Table 8-2 for the pin
configuration.
Table 8-2. Pin configuration.
Status signal patterns
Table 8-3. Status signals for pin 14 - 18 (see “Pin configuration.” on page 31).
Pin
Type
Signal name
1
-
GND
2
-
GND
3
-
GND
4
IN
IN_0
5
IN
IN_1
6
IN
IN_2
7
IN
IN_3
8
IN
IN_4
9
OUT
CAL_CONF
10
OUT
OUT_6
11
-
GND
12
-
GND
13
-
GND
14
OUT
DET_ERROR
15
OUT
LEAK_OUT
16
OUT
DET_ON
17
OUT
DET_SIGNAL
18
OUT
DET_WAIT
19
OUT
OUT_0
20
OUT
OUT_1
21
OUT
OUT_2
22
OUT
OUT_3
23
OUT
OUT_4
24
OUT
OUT_5
25
OUT
24 VDC OUT
Signal
Function
DET_SIGNAL
Gas detected / Sensor not recovered.
DET_WAIT
High during warm-up.
DET_ON
High when detector is on.
LEAK_ALARM
Leak above Reject Level detected.
DET_ERROR
High if Probe, Sensor or Cable is broken.