En probe control port connector, Status signal patterns, Signal function – INFICON Sensistor ISH2000P Hydrogen Leak Detector, Panel Model User Manual

Page 33

Advertising
background image

Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions


OP-Sensistor ISH2000-EN-201203

31

EN

Probe Control Port Connector

The control port connector is a 25-pin female D-sub. Refer to Table 8-2 for the pin
configuration.

Table 8-2. Pin configuration.

Status signal patterns

Table 8-3. Status signals for pin 14 - 18 (see “Pin configuration.” on page 31).

Pin

Type

Signal name

1

-

GND

2

-

GND

3

-

GND

4

IN

IN_0

5

IN

IN_1

6

IN

IN_2

7

IN

IN_3

8

IN

IN_4

9

OUT

CAL_CONF

10

OUT

OUT_6

11

-

GND

12

-

GND

13

-

GND

14

OUT

DET_ERROR

15

OUT

LEAK_OUT

16

OUT

DET_ON

17

OUT

DET_SIGNAL

18

OUT

DET_WAIT

19

OUT

OUT_0

20

OUT

OUT_1

21

OUT

OUT_2

22

OUT

OUT_3

23

OUT

OUT_4

24

OUT

OUT_5

25

OUT

24 VDC OUT

Signal

Function

DET_SIGNAL

Gas detected / Sensor not recovered.

DET_WAIT

High during warm-up.

DET_ON

High when detector is on.

LEAK_ALARM

Leak above Reject Level detected.

DET_ERROR

High if Probe, Sensor or Cable is broken.