Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions

32

OP-Sensistor ISH2000-EN-201203

EN

DET_ERROR will go high for a short time (1-5 seconds) when the detector is switched
on. It will go low when the sensor has been checked.

In normal operation, DET_ERROR = HIGH means that there is a problem with the
sensor, probe, or cable.

DET_WAIT is high when instrument is in warm-up mode after switching on power.
Instrument will also go into warm-up if there is a temporary fault in the sensor or
sensor connection.

The timing of the status signals in relation to different events is described by the
following two examples:

Example:

Input signals issued to control the APC system should have a pulse length

of at least 40 ms.

Example:

Output signals switch with a cycle time of 20 ms (0.02 s). This is the cycle

time of the APC system.

Note:

Not valid for battery operated version of Sensistor ISH2000.

After power on

Fig 8-3. Status signals after power on.

When detecting a gas signal

Fig 8-4. Status signals when detecting a gas signal.

Power on

Power off

Warming up

Detection or analysis

mode. No gas detected.

DET_SIGNAL

LEAK_OUT

DET_WAIT

DET_ON

DET_ERROR

Gas signal

Purge level

Detector signal level

0 PPM

DET_SIGNAL

LEAK_OUT

PURGE_LEVEL Trig-

ger (APC)

Purge level

Reject level