Hydrogen leak detector Sensistor ISH2000

-

Operating Instructions


OP-Sensistor ISH2000-EN-201203

37

EN

The active test cycle defined by the APC driver starts. “M” is returned if the selected
driver supports active test. “F” (failed) is returned otherwise.

N = Print Request

Returns current analysis value.

Q = Set APC in stand by, (stop a measurement)

Returns a “Q”.

R = Activates probe 0 (built in P50 driver)

Returns an “R”.

S = Activates probe 1 (Installed probe driver)

Returns an “S”.

X = Shift State to "Analysis Mode"

Returns nothing.

Z = Shift State to "Detection Mode"

Returns nothing.

Y = Shift State to “Combined Mode”

Return nothing.

Supported parameters

Table 8-12. The following parameters can be downloaded to Sensistor ISH2000 in

Analysis

and

Detection mode

.

* entered as integer in 10’s of seconds, 1= 0.1s, 100 = 10s, 60000= 6000s

Transfer of parameters

Send parameters one by one; first send the specific header (for example “A”), then
wait until the RS232 transmitter unit has sent the data (approximately 20 ms), and
then send the data (for example “1.00E+01”). Data string must be ended with a
carriage return character, chr13 (dec).

Example:

“CPPM” or “C PPM”, Carriage Return (chr 13). This sets Reject Level Rate

Unit to “PPM”.

Parameter

Header

Data

Reject Level

A

n.nnE+nn

Correlation Value

B

n.nnE+nn

Analysis Unit

C

Text string (max 12 characters)

Analysis Unit

CUx

x=1 to 8, 1=PPM, 2=CC/S up to 8

Timer A

D

nnn*

Timer B

E

nnn*

Timer C

F

nnn*

Timer D

G

nnn*

Purge Level

H

n.nnE+nn

Reference Value

I

n.nnE+nn

Reference Unit

J

Text string (max 12 characters)

Reference Unit (Program settings)

JUx

x=1 to 8, 1=PPM, 2=CC/S