Chapter 3 solenoid valve assembly installation, 1 introduction, 2 installation with the uhv bakeable sensor – INFICON UHV Bakeable Sensor User Manual

Page 31: Chapter 3, Oid valve (see

Advertising
background image

3 - 1

PN

07

4-

15

4N

UHV Bakeable Sensor Operating Manual

Chapter 3

Solenoid Valve Assembly Installation

3.1 Introduction

The solenoid valve assembly (PN 750-420-G1, see

Figure 3-2 on page 3-4

) and

the shuttered UHV Bakeable Sensor assembly should be installed at the same
time.

3.2 Installation with the UHV Bakeable Sensor

1

Align the score line on the solenoid valve bracket (see

Figure 3-2 on page 3-4

)

over the edge of a table or other square edge.

2

Using pliers, grasp the part of the bracket extending over the edge and push
down. The assembly will break along the score line.

3

Use a file to smooth any rough edges which occur along the break.

4

Insert the bakeable sensor through the mating flange on the vacuum chamber.

5

Place the solenoid valve bracket at the desired position on the outside of the
feedthrough, with the two holes in the bracket aligned with two of the
feedthrough bolt holes.

6

Install and tighten the feedthrough bolts.

7

Remove the quick disconnect air fitting from the exhaust port of the
solenoid valve (see

Pneumatic Connections, section 3.3

) and thread it into the

#10-32 fitting adapter on the feedthrough air tube.

8

Connect the 3.175 mm (0.125 in.) air tube from the A port of the solenoid valve
to the quick disconnect fitting installed in step 7 (see

Pneumatic Connections,

section 3.3

).

Advertising