5.1.2.4 XIU Test Function

The XIU Test function is a feature of IC/5, Cygnus, IC6, Cygnus 2, and XTC/3
controllers. Refer to the controller operating manual for instructions on using
the XIU test function.

5.1.3 Digital Multimeter

A useful tool for diagnosing sensor problems is the Digital Multimeter (DMM).

To isolate the cause of a sensor problem, perform electrical isolation and continuity
checks, starting with the Electrical Isolation Check (

section 5.1.3.1

).

5.1.3.1 Electrical Isolation Check

1

Remove the crystal holder from the bakeable sensor.

2

Disconnect the BNC cable from the bakeable sensor feedthrough.

3

Select the DMM ohmmeter function and a high megohm (M

) resistance scale.

4

At the BNC connector on the feedthrough, measure the resistance between the
center contact and shield, as shown by

Figure 5-1

.

If resistance is more than 10 M

, electrical isolation is good. Go to

section

5.1.3.2, Electrical Continuity Check, on page 5-6

.

If resistance is less than 10 M

, contact INFICON (refer to

section 1.3 on

page 1-2

).

Figure 5-1 BNC connector resistance