3 pneumatic connections, Pneumatic connections, section 3.3, And thre – INFICON UHV Bakeable Sensor User Manual

Page 32: Pneumatic connections, Section 3.3

Advertising
background image

3 - 2

PN

07

4-

15

4N

UHV Bakeable Sensor Operating Manual

9

Attach the P port of the solenoid valve to a source of air. The air supply must
be 70 psi (gauge) {85 psi (absolute)} (5.8 bar (absolute)) [584 kPa (absolute)]
(minimum) to 80 psi (gauge) {95 psi (absolute)} (6.5 bar (absolute))
[653 kPa (absolute)] (maximum) (see

Pneumatic Connections, section 3.3

.)

WARNING

Do not exceed 100 psi (gauge) {115 psi (absolute)}
(7.9 bar (absolute)) [791 kPa (absolute)].

Connection to excessive pressure may result in personal
injury or equipment damage.

CAUTION

Maximum temperature for the solenoid valve assembly
is 105 °C for bakeout and operation.

10

Make electrical connections to the solenoid valve (see

Electrical Connections,

section 3.4

).

3.3 Pneumatic Connections

Figure 3-1 Pneumatic solenoid valve tube connections

To Air
Fitting Of
Feedthrough

(Normally Open)

A Output

Port

P Supply

(Normally Closed)

Tube Fitting

Air Supply

(Provided With Valve)

Exhaust