6 Safety Manual
59
-
Damage and corrosion of temperature probes – protection tubes
-
Corrosion and correct positioning of the contacts and terminals of cable connections
-
Seals of terminal heads and cable ducts
-
Interruptions due to "knocking" on the temperature probe/measuring insert
Since the maximum operating temperature influences the drift behavior, the temperature probe should be recalibrated or re-
placed at certain intervals to ensure reliable and precise temperature measurement.
The testing intervals are listed in the table below:
6.17 Intrinsic safety according to DIN EN 60079-11
v
Table 16: Testing intervals
Maximum operating temper-
ature
Pt - RTD temperature probes
Thermocouples
200 °C
5 years
5 years
550 °C
2 years
5 years
700 °C
1 year
2 years
1000 °C
Base metal 1 year
Noble metal 2 years
1500 °C
1 year
H
The testing intervals specified here are recommendations that must be adapted to the special conditions at the op-
erating location and,
if necessary, the user should perform the tests more regularly.