Leica RM2125 RTS

4. Startup

The specimen clamps (standard or

universal cassette clamps) can also

be fastened directly to the specimen

holder fixture.

4.5 Direct mounting of the specimen clamp on the specimen holder fixture

• Attach the specimen clamp (12) to the speci-

men holder fixture as shown (

Fig. 9.3

) and

fasten it using the two screws (17).

• To remove the dovetail guide (13) from the

specimen clamp (12), unscrew the two

screws (17) from the dovetail guide (

Fig. 9.2

).

Again, use the Allen key No . 3.

• Then remove the dovetail receptacle (14)

from the specimen holder fixture (

Fig. 9.1

).

To do so, loosen and unscrew the four

screws (16) using an Allen key No. 3 (18).

• Move the specimen holder fixture (10) to the

upper end position by turning the handwheel

and engage the handwheel lock.

To do so, proceed as follows:

Fig. 9

17

16

14

10

18

12

12

13