42

Instructions for Use, V2.1 – 10/2012

6. Cleaning and Maintenance

Knife holder E

Fig. 35

Take the knife holder apart for cleaning. The

front pressure plate can be removed for clean-

ing. To do so, proceed as follows:
• Fold knife guard (5) downward.

• Rotate the blade clamping lever (89) down-

ward.

• Carefully remove the blade and dispose of it

properly.

• Pull out the clamping lever (89) sideways.

• Remove pressure plate (86).

• Clean all parts of the knife holder.

If several knife holders are cleaned at the

same time, the parts must NOT be mixed

up! Failure to adhere to this may result in

sectioning problems!

For cleaning and removal of paraffin, do

not use xylene or cleaning fluids contain-

ing alcohol (e.g. glass cleaner).

• Lay the removed parts on an absorbent

cloth into the drying chamber (up to a max.

of 65 °C) and let the paraffin contamination

run off.

There is a burn hazard when removing

the parts from the drying chamber (65°C).

Wearing safety gloves is recommended!

• After cleaning the moving parts, apply a

thin coat of drive part oil to them (

see also

Chapter 6.2

).

5

5

89

86

89

80

86