Operation – Leica Biosystems RM2125 RTS User Manual

Page 25

Advertising
background image

25

Leica RM2125 RTS

The Leica RM2125 RTS is equipped with a me-

chanical trimming function that is activated via

the trimming lever (35).
The trimming lever has 3 click stops:

0 µm, 10 µm, and 50 µm.
The points (36) mark the two trimming stages:

= 10µm



= 50 µm

• To activate the trimming function, press the

lever downwards into the desired position

and keep it pressed down.

• With each rotation of the handwheel, a feed

motion of 10µm or 50 µm takes place.

• After you let go of the lever, it springs back

to its original position (zero position). The

trimming function is thereby deactivated.

The section thickness that has been

set is not added to the selected trim-

ming value.

If the section thickness that has been

set is greater than the selected trim-

ming value, the section thickness is

fed.

• Bring the specimen closer to the knife by

rotating the coarse feed wheel.

• Select the desired trimming stage.
• Trim the specimen by rotating the hand-

wheel (15) until the desired specimen plane

is reached.

• Release the trimming lever (35).

Trimming with the mechanical trimming function

15

35

35

36

Fig. 17

5. Operation

Advertising