20

Instructions for Use, V2.1 – 10/2012

5. Operation

• The index marks (0°, 5° and 10°) for adjust-

ment of the clearance angle (27) are located

on the right side of the knife holder (25).

• There is also an index mark (28) on the right

side of the knife holder base (21) which

serves as a reference point when adjusting

the clearance angle.

• When using the knife holder base (non-

orientable), loosen the Allen screw to

release the clamp by turning Allen key No. 4

(24) counterclockwise.

• When using the knife holder base (with

lateral movement), rotate the lever on the

right side of the knife holder base counter-

clockwise.

• Move the knife holder until the index mark of

the desired clearance angle coincides with

the reference line on the knife holder base.

Example:
Enlarged detail showing a clearance angle

setting of 5°.

The recommended clearance angle

setting for knife holder E is approx.

1° - 3°.

• Firmly hold the knife holder in this posi-

tion and rotate the lever (24) or Allen screw

(depending on the knife holder base used)

clockwise to clamp it.

5.3 Adjusting the clearance angle

Fig. 12

21

24

25

27

28

Enlarged detail:

Index marks for

clearance angle

setting